Incubator Bayi YP-100
Overview
Specifications
Packaging

Power requirements AC220V-230/AC110V-120V, 50/60Hz
Power requirements 680VA maximum
Power outlet AC220V-230V, 50/60Hz,Current 2A max
 AC110V-AC120V,50/60Hz,Current 2Amax
Temperature control range and correlative
Air mode control 25-37°C
Temperature rise time (typical @22°C ambient) < 45minutes
Temperature variability < 0.5°C
Temperature uniformity (level mattress) <0.8°C
Alarm
Power alarm power failure
Motor alarm  motor failure
 Motor run speed too low
 Air flow stop
Sensor alarm  Air temperature sensor failure
 Isolate overheating sensor failure
 Air flow sensor failed
Temperature deviation alarm High/low air temperature
Over temperature alarm Air temperature control mode
 Activated if incubator temperature
 not exceeds 38°C
E² ROM failed  internal malfunction
Environment temperature (not to use in the environment exceed specified)
Operating range  +20-+30°C
Storage range -40-+70°C
Environment humidity
Operating range 30%RH-75%RH non-condensing
Storage range  < 80% RH non-condensing
Air Velocity Ambient Air Velocity  < 0.3 m/s
Atmospheric Pressure
Shipment & store atmospheric pressure range 500hPa-1060hPa
Operating atmospheric pressure range 700hPa-1060hPa
Physical Dimensions
Height from mattress to floor 90cm
Width 100cm
Depth 150cm
Incubator main body weight > 60kg
Stand weight > 10kg
I.V Pole max. Load 20N
Hood max. Load 4kg
Infant mattress max. Load 10kg
Other specification
Infant mattress width 31cm
Infant mattress length 63cm
Noise level within hood environment <60dB (A)
 (Base on surroundung environment noise below 45dB)
Carbon diaxide(Co2) level within the hood less than 0.5% when a mixture of Co2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress.
Air velocity over mattress  less than 0.35m/s
Reset Key
When the following alarm occurs, press RESET/SILENT key to silence the alarm for 5 minutes, the alarm will come back automatically after 5 minutes. If alarm condition no longer exists, the alarm will be canceled (except overheat); or press the RESET/SILENT key twice to cancel the alarm, the equipment will come back to set state
Temperature deviation alarm
Sensor alarm
Motor alarm
Over temperature

Overview
Specifications
Packaging
Overview
Specifications
Packaging
Similar Items
Latest News
Main Nav
Home
About Us
News
Articles
Career
Video
Contact Us
Sitemap
Products
Antiseptic & Desinfectan
OneMed Diagnostic
Medical Disposable
OneLab
Wound Care
Elektromedik
Dental Unit
Hospital Furniture
OneMed GLOVE
Oxycare
Urology (Bioteq)
Tag Clouds
Medical Antiseptic Hospital Health Disposable Desinfectant Sterile Aseptic Infeksi Diagnostic Care
Get in touch with us
OneMed BUILDING
Jl. Mayjend Sungkono, Komplek Darmo Park Blok IV No. 14-15
Surabaya 60225 - Jawa Timur
 
PHONE
+62 31 5672 236
FAX
+62 31 5688 912
 
EMAIL
sales@onemed.co.id
Find Us

show/hide

© 2011-2018 OneMed Health Care. All rights reserved.